OE2031 Lock-in Amplifier
DC-10 MHz Digital Lock-in Amplifier
DC to 10MHz frequency
1 nV to 2 Vrms full-scale sensitivity
3nV/√Hz input noise
Time constants from 100 ns to 3 ks
>120 dB dynamic reserve
6 demodulators and 2 PLLs
Oscilloscope, Spectrum Analysis, Sweeper,
4 PID Controllers, AM/FM/PM-MOD
Introduction
性能参数
Performance
Application Area
Software Interface
资料下载
相关视频

Overview

The OE2031 is a high-frequency phase-locked amplifier offering exceptional measurement capabilities for both weak signals and wide bandwidths, with a bandwidth measurement range of DC to 10MHz. With the advantage of the latest digital signal processing technology and high-speed 250MSPS 14-bit ADC, OE2031 can easily detect the phase and the magnitude of weak signals overwhelmed by various large noise. The performance of OE2031 is as good as other lock-in amplifiers all over the world, even better than them in some certain parameters, such as measurement accuracy, SNR, dynamic reserve, which meets the needs of scientific research and industrial application well.


Input Channel

OE2031 detects an input signal in a single-ended mode or a differential voltage mode. With an ultra low-noise pre- amplifier, the input noise is as low as 3 nV/Hz@1MHz. The input impedance is 50Ω or 10MΩ and the full-scale input voltage sensitivity ranges from 1nV to 2 Vrms. Besides, designed to eliminate power frequency interference. A programmable gain amplifier is used to adjust the dynamic reserve of the system, so that OE2031 can keep a high dynamic reserve of 120 dB. The high-precision 14-bit ADC has a sampling rate of 250 MSPS, and the excellent anti- aliasing filter (Blocking Attenuation >100dB) in front of the ADC can effectively prevent signal aliasing.


Reference Channel

The reference signal can work in external mode or internal mode. In internal mode, a precise and stable internal oscillator generates sine wave as an internal reference that is multiplied by the input signal. This internal signal is without any phase noise. With the digital phase-shifting technique, the phase resolution of the reference signal is 0.001deg. OE2031 can work at any fixed frequency from DC to 10MHz in this mode.In external mode, the reference signal can be a sine wave or a TTL pulse or square wave. The rising or falling edge of the external reference signal triggers the Phase Lock Loop (PLL) to lock the external signal. Based on the frequency of the reference signal, can demodulate multiple harmonics and arbitrarily frequency input signal. The maximum harmonic signal frequency can reach 10000 times the fundamental frequency, and the maximum harmonic frequency cannot exceed the maximum operating frequency of the instrument by 10MHz.


Digital Demodulator and Output Filter

The key component of the OE2031 is the digital demodulator. Compared to traditional analog lock-in amplifiers, OE2031's internal digital demodulator effectively rejects the measurement errors caused by DC drift and offset. In addition, by optimizing the multiplication of the internal coherent signal of the digital demodulator, the calculation error is minimized so that the instrument can accurately detect the input weak signal. OE2031 features 6 dual-phase demodulators. Time constants of the output low-pass filter from 100ns to 3ks (50μHz to 1.6MHz) can be selected with a choice of 6, 12, 18, 24, 30, 36, 42 and 48 dB/oct rolloff (1st to 8th order). This low-pass digital filter is implemented using a high performance digital filter with a sample rate of 250 MHz . The digital demodulation and the low-pass filter used in OE2031 guarantees a high dynamic reserve (>120dB), accurate phase (absolute phase error < 1deg). Moreover, when the frequency of the input signal is lower than 1MHz, A synchronous filter can be used to eliminate the harmonic influence of the reference signal, ensuring that OE2031 can detect a low- frequency signal quickly and effectively.


Auxiliary IO

The OE2031 features arbitrary auxiliary input and output interfaces. The AUX-IN port can measure voltages below 10V at a sampling rate of 312.5kSPS. The AUX-OUT/CH-OUT port operates at a refresh rate of 1 Msps, capable of outputting X, Y, R and θ-values, and arbitrary DC voltages. Additionally, the OE2031 incorporates CLK-IN, CLK-OUT, TRIG IN, TRIG OUT, and analogue mirror output ports.


Signal Generator

The OE2031 utilises a 16-bit high-precision digital-to-analogue converter (DAC) to output a sinusoidal signal matching the internal reference frequency. This sinusoidal wave features a maximum amplitude of 2Vrms, an accuracy of 0.1μVrms, and a frequency range spanning DC to 10MHz. The amplitude and phase of the output sine wave can be configured via the OE2031 display.


Internal Oscillator

The two internal oscillators of the OE2031 generate low-distortion (-75dBFS) sine reference signals from DC to 10 MHz/80 MHz, with frequency resolution up to 1 nHz. The frequency and amplitude of the reference signal can be set via the OE2031 front panel or communication interface. When the OE2031 is configured in external reference mode, the internal reference signal locks to the external reference signal.


Display

The OE2031 is equipped with a 5.6-inch TFT-LCD colour display serving as the primary interface for the instrument, enabling comprehensive independent control via the keyboard. Data detectable by the instrument, such as signal X, Y, R and θ-values, can be displayed through numerical graphs, bar charts, and full data presentation modes.

彩色


Manual Operation

The OE2031 can use the softkeys on the front panel, together with the knobs, to realize convenient and quick parameter adjustments, such as adjusting the frequency and phase of the internal reference signal.


Auto Function

The OE2031 can automatically adjust to different optimal operating states based on the input signal, such as automatic range adjustment, automatic filter parameter setting, automatic phase compensation adjustment, and automatic output bias adjustment.


Communication Interface

The OE2031 has USB2.0, Gigabit Ethernet port and standard 9-pin RS232 serial port. Through these interfaces, users can effectively utilize all testing functions of the OE2031 on a controlling computer, setting reasonable control parameters and reading the data measured by the instrument.


aa
111
bb
222
cc
333
Signal Channel

Voltage Input Mode                                               Single-ended or Differential

Full-scale Sensitivity                                             1 nV to 1 V in a 1-2-5 sequence

Impedance

      Voltage                                                            50 Ω // 5pF or 10 MΩ // 5pF,AC or DC coupled

C.M.R.R                                                                >70 dB to 100 Hz

                                                                              >50 dB to 100 kHz

Dynamic reserve                                                   >120 dB

Gain accuracy                                                       0.5% typ(<1MHz), 3% max

Voltage Noise                                                        3nV/√Hz at 999.99 kHz

Line filters                                                              50/60 Hz and 100/120 Hz

Gounding                                                               BNC shield can be grounded or floated via 10 kΩ to ground


Reference Channel

Input

      Frequency range                                             10 uHz to 10 MHz

      Reference input                                               TTL or Sine

      Input impedance                                              1 MΩ

      Square reference level                                     VIH>3V, VIL<0.5V

      Sine reference signal                                        >1 Hz

                                                                                > 400 mVpp

Phase

      Resolution                                                        1udeg

      Absolute phase error                                        <1°

      Relative phase error                                         <1 mdeg

      Phase noise                                                     Internal ref. Synthesized, <0.0001 deg at1 kHz

                                                                               External ref. 0.001 deg at 1 kHz (100 ms time constant, 12 dB/oct)

Drift                                                                        <0.01 deg/℃ below 100 kHz

                                                                               <0.1 deg/℃ above 100 kHz

Harmonic detection                                                2F, 3F, …nF to 10 MHz (n<32,767)

Acquisition time                                                     Internal Ref. Instantaneous acquisition

                                                                              External Ref. (2 cycles + 5 ms) or 40 ms,whichever is larger


Demodulator

Number of demodulators                                       4

Stability

      Digital outputs                                                  no zero drift on all setting

      Display                                                             no zero drift on all setting

      Analog outputs                                                 <5 ppm/℃ for all dynamic reserve settings

Harmonic rejection                                                 -90 dB

Time constants                                                       30 ns to 4.4 ks (6,12,18,24,30,36,42,48 dB/oct rolloff)

Synchronous filters                                                 Available below 200 Hz(18,24,30,36,42,48 dB/oct rolloff)


Internal Oscillator

Frequency                                                             10 uHz to 10 MHz

      Accuracy                                                          2 ppm + 10 µHz

      Resolution                                                        1 nHz

      Disrortion                                                         -80 dBc (f<10 kHz),-60 dBc (f>1 MHz)

Amplitude                                                               1 uVrms to 1 Vrms ( Resolution:1 uVrms)

      Accuracy                                                          0.5%

      Stability                                                            50 ppm/℃

Sine Outputs                                                          Sine signal,output impedance 50 Ω

TTL Outputs                                                           5V TTL/CMOS level,output impedance 50Ω


AUX Inputs and Outputs

AUX Inputs

      Function                                                           4 Channel Inputs

      Amplitude                                                         ±10 V,1 mV resolution

      Impedance                                                       1 MΩ

AUX/CH Outputs

      Function                                                           4 Channel Outputs

      Amplitude                                                         ±10 V,0.1 mV resolution

      Drive current                                                    30mA max


Display

Screen                                                                   5.6 inch 640×480 TFT

Screen format                                                        Single or dual display

Display quantities                                                   Each display shows one trace,

                                                                               traces can be defined as X,Y,R,θ

Display types                                                          Numerical form, bar graph


Interfaces

USB2.0

RS-232(DB-9)

1000Mbps Enthernet


General 

Power requirements

      Voltage                                                             220-240 V AC

                                                                               100~120 VAC(optional)

      Frequency                                                        50/60 Hz

      Power                                                               50 W

Power supply rejection                                           70dB@1MHz

Weight                                                                    12 KG

Dimensions

      Width                                                                448 mm

      Depth                                                                513 mm

      Height

            With feet                                                     148 mm

Scanning Microscope                                                AFM, STM, SPM

Materials Science                                                       Carrier mobility, Carrier density, Hall effect, Ultrasonic materials  

Transport Measurement                                            Conductivity measurement, Impedance measurement

Noise Represents                                                       Noise density, Cross-correlation measurement

Optical Experiment                                                    Spectral analysis, Spectral measurement, THz measurement, TDLAS 

Sensor Measuring                                                      Gyroscope, Photoelectric sensor, Resonator, Accelerometer

Magnetic Sensor                                                        SQUIDs, NV color center, Atomic Magnetometer, VSM

Biomedical                                                                  Microfluidic

Lock-in Amplifier

锁放界面


Parametric Sweeper


PID Controllers


FFT Spectrum Analyzer


Oscilloscope

示波器

文件名称
说明
大小
版本号
下载
OE2031介绍文档
OE2031锁相放大器产品介绍文档
2.17MB
v251126
OE2031用户使用手册
OE2031锁相放大器使用操作说明书
18.45MB
v260617
SSI Lucid Suite 控制平台软件
电脑端控制平台软件,适用于OE2031、OE2041锁相放大器
47.52MB
v260715
串口调试助手
串口调试助手,方便对仪器进行调试
485.02KB
v3.8.3
申请试用