OE1022 Lock-in Amplifier
102kHz Digital Lock-in Amplifier
1mHz to 102kHz frequency 
range10us to 3ks time constants
>100 dB dynamic reserve1nV to 
1nV to1V full-scale sensitivity
Multiple-harmonic measurement 
FFT spectral analysis
5.6-inch color display
Introduction
Performance
Application Area
Software Interface
Data download
Related videos

Input Channel

    With the low-noise analog pre-amplifier, the signal input of OE1022 can be switched to operate in the single-ended or differential voltage mode, and the input noise is 10 nV/√Hz. The input impedance is 10 MΩ and the full-scale input voltage sensitivity ranges from 1 nV to 1V. Besides, OE1022 can also be used for current measurements with variable current gains of 10^6 or 10^8 V/A. Two line filters (50/60 Hz and 100/120 Hz) are designed to eliminate line related interference. The programmable gain amplifier is provided to adjust the dynamic reserve of the system according to the magnitude of the input signal, so that OE1022 has inherently large dynamic reserve up to 100dB. The sampling rate of 312.5KSPS is determined by a precision 24-bit A/D converter and a specific filter is designed to avoid aliasing.


Reference Channel

    In order to provide the reference signal for OE1022, an externally applied sine wave or square wave, or its owninternally synthesized reference source could be used. When the instrument is set to internal reference mode, the internal precision stabilized oscillator and the digital synthesized algorithm are used to generate sinewave output that multiplies the input signal, there is virtually no reference phase noise when choosing internal reference mode. Taking advantage of digital phase-shifting technique, the reference signal phase could be adjusted with 0.01° resolution. The internal reference mode can operate at a fixed frequency from 1 mHz to 100 kHz. In addition, the external reference is also applicable to OE1022, including the sinewave reference signal and TTL logic reference signal. The rising and trailing edge of external reference signal are applied to trigger off the internal Phase Locking Loop (PLL). Based on the frequency of reference signal, harmonic detection can be performed by OE1022. The maximum frequency of the measurable harmonics is 32767 times of basic frequency, and it is also less than the maximum operational frequency 100 kHz.


Display

    OE1022 uses the 5.6 inch 640×480 TFT color display as its screen. Data measured by OE1022, such as, X, Y, R, θ, is stored in up to four traces. Trace values can be displayed as a bar graph or as a strip chart showing the trace values as a function of time.

image

    Besides, OE1022 can display polar plots, showing phasor composed by in-phase and quadrature components of the signal. All displays can be easily scaled by the manual operation, and the auto-scale feature is available to optimize display quickly. The screen can be configured as a single large display, or two horizontally-split displays.

image


image


Simultaneous Multiple-harmonic Measurement

    The traditional phase-locked amplifier can only measure fundamental frequency signal or one certain harmonic ware at the same time, so it can not meet the requirement in some occasions when measuring multiple frequency components simultaneously is needed. In digital end, OE1022 combined FPGA and ARM technology, which has wider processing bandwidth and more flexible digital framework. The digital processing precision could reach 48 bits and could measure 3-channal harmonic components simultaneously, which makes one OE1022 is equal to three traditional phase-locked amplifiers.

 

Remote Operation

    The built-in RS232 to USB interfaces on OE1022 allow full manual operation from a controlling computer, including setting or interrogating control and reading out data. Free LabVIEW program is provided. It makes it easy to set up and run complex experiments, such as remote control of every instrument function. The display menu offered by the LabVIEW program allows customers observe all command received and responses generated by instrument.



Signal Channel

        Voltage input Mode                   Single-ended or Differential

        Full-scale Sensitivity                  1 nV to 1 V in a 1-2-5 sequence

                                                           1 fA to 1 µA

        Current input                             106 or 108 V/A
        Impedance
                Voltage                              10 MΩ
                Current                              1 kΩ to virtual ground
        C.M.R.R                                     >100 dB to 10 kHz, decreasing
        Dynamic reserve                        >120 dB
        Gain accuracy                            0.2% typ, 1% max

        Voltage Noise
                                                           5 nV/Hz at 997 Hz
        Current Noise
                                                           5 fA/√Hz at 97 Hz
   
                                                       13 fA/√Hz at 997 Hz
        Line filters                                   50/60 Hz and 100/120 Hz
        Gounding                                     BNC shield can be grounded or floated via 10 kΩ to ground


Reference Channel


        Input
                Frequency range                1 mHz to 102 kHz
                Reference input                  TTL or Sine
                Input impedance                 1 MΩ
                Square reference level        VIH>3V, VIL<0.5V
                
Sine reference signal          >1 Hz
                                                            > 400 mVpp
        Phase
                 Resolution                          0.001°
       
         Absolute phase error          <1°
                 Relative phase error           <1 mdeg
                 Phase noise
                                                            Internal ref. Synthesized, <0.0001 deg at1 kHz
                                                            External ref. 0.001 deg at 1 kHz (100 ms time constant, 12 dB/oct)
                 Drift
                                                            <0.01 deg/℃ below 10 kHz
                                                            <0.1 deg/℃ above 10 kHz

        Harmonic detection                     2F, 3F, …nF to 102 kHz (n<32,767)

       

        Acquisition time

                                                            Internal Ref.     Instantaneous acquisition

                                                            External Ref.        (2 cycles + 5 ms) or 40 ms,whichever is larger


Demodulator     

        Stability            

                Digital outputs                      no zero drift on all setting

                Display                                 no zero drift on all setting

                Analog outputs                     <5 ppm/℃ for all dynamic reserve settings

        Harmonic rejection                     -90 dB

        Time constants                           10 µs to 3 ks (<200 Hz)

                                                           10 µs to 30 s (>200 Hz)

        Synchronous filters                     Available below 200 Hz(18, 24 dB/oct rolloff)


Internal Oscillator  

        Frequency                                 Range 1 mHz to 102 kHz

         Accuracy                                   2 ppm + 10 µHz

                Resolution                         1 mHz

        Disrortion                                   -80 dBc (f<10 kHz),-70 dBc (f>10 kHz)

        Amplitude                                   0.001Vrms to 5 Vrms ( Resolution:1 mVrms)

                Accuracy                            1%

                Stability                               50 ppm/℃

        Sine Outputs

                                                           Sine signal,output impedance 50 Ω

        TTL Outputs                               5V TTL/CMOS level,utput impedance 200Ω


Display

        Screen                                       5.6 inch, 640×480 TFT

        Screen format                            Single or dual display

        Display quantities                       Each display shows one trace,

                                                            traces can be defined as X,Y,R,θ

        Display types                              Numerical form, bar graph, polar plot and strip chart

        



AUX Inputs and Outputs  

        CH1 and CH2 Outputs

                Function                             Output X, Y, R, θ

                Output Voltage                    ±10 V full scale.

                                                            30 mA max output current

                Update Rate                        312.5kHz

        AUX Inputs

                Function                             4 Channel Inputs

                Amplitude                           ±10 V,1 mV resolution ratio

                Impedance                         1 MΩ

        AUX Outputs

                Function                             4 Channel Outputs

                Amplitude                           ±10 V,1 mV resolution ratio

                Drive current                      ±25mA max

        Trigger Input

                Function                             TTL external trigger is used for data storage

        Monitor Output

                Function                             Analog output of a signal-amplifier

                Drive current                      ±40mA max


Interfaces

        RS-232 to USB interface,

        IEEE-488 interface(optional).


General

        Power requirements

                Voltage                                220~240 V AC

                                                            100~120 VAC(optional)

                Frequency                           50/60 Hz
                Power                                  30 W

        Power supply rejection                70dB@1MHz

        Weight                                         11 KG      

        Dimensions

                Width                                  448 mm

                Depth                                  513 mm

                      

                Height

                        With feet                     148 mm



Scanning microscope                AFM、STM、SPM

Materials science                      Carrier mobility, Carrier density, Hall effect, Ultrasonic materials  

Transport measurement            Conductivity measurement, Impedance measurement

Noise represents                       Noise density, Cross-correlation measurement

Optical experiment                    Spectral analysis, Spectral measurement, THz measurement, TDLAS 

Sensor measuring                     Gyroscope, Photoelectric sensor, Resonator, Accelerometer

Magnetic Sensor                       SQUIDs, NV color center, Atomic Magnetometer, VSM

Biomedical                                Microfluidic  

Chinese Interface

OE1022中文软件界面



English Interface

OE1022英文软件界面

File name
Explain
Size
Version number
Download
OE1022_Introduction
OE1022 102kHz LIA's introduction
1739kb
v20210804
OE1022_Console_EN
OE1022 102kHz LIA LabVIEW Console
1156kb
v20210903
UartAssist
Serial debugging assistant, easy to debug the instrument
486kb
v3.8.3
Uart Driver
Windows 7/8 Users need to install the driver before using the USB serial port and RS232 serial port
4316kb
v2.12.26
USB_Driver for Win7/8
Users of windows 7/8's need to install this USB2.0 driver
24820kb
v1.5.0
OE1022 LabVIEW Library
LabVIEW library functions for OE1022, only support LabVIEW2011 and above
451kb
v20170509
OE1022_User_Manual_en
OE1022 102kHz LIA's user manual
7050kb
v20180402
Apply for probation